Product model
BT100-1F Dispensing type Peristaltic Pump
[Product introduction]
Flow range: 0.0002-48ml/min
The best filling volume: 0.1-100mL
Various pump head for optional, DG series, YZ series、BZ series
Working mode: Flow mode, Dispensing mode
Flow rate rectify, Dispensing volume rectify function
Flow accuracy within ±0.4%
[Application]
Lab, filling
Reagent packing, culture medium filling
[Technical parameter]
Model |
BT100-1F Dispensing type Peristaltic Pump |
CPU |
ARM Cortex-0 Core,STM32 Chip |
Motor drive chip |
SANYO THB7128 drive chip |
Speed resolution |
0.1RPM |
Distribute function |
liquid Quantitative filling accurately |
Dispensing type range |
0.1mL-999L |
Filling times |
1-9999 times |
Interval time |
0.1S-24Hr |
Display function |
128×32 Dot matrix LCD display, flow rate and speed display in the same screen |
Direction control |
Clock-wise/counter clock-wise direction reversible |
Operation mode |
Membrane keypad adjust the speed, also support external signal adjust speed |
Full speed function |
Full speed button to realize fast filling and drain |
Power-down memory function |
After power on, running as before |
Back suck function |
Running under Dispensing type mode, pump reverse a certain angle after stopping, to prevent liquid drip |
Calibration function |
Flow rate rectify function, accuracy ±0.5%
|
Dispensing type Calibration function |
Input filling volume , it rectify the distributed liquid automatically; accuracy ±0.5%
|
External control function |
Different model external control module for Start/Stop, direction, flow rate; 0-5V、0-10V、4-20mA、0-10KHz 4 module for option |
Communication function |
Optional communication module, RS485 communication control (via Modbus protocol) |
Foot switch control |
Optional |
Bracket |
Optional (for installing the nozzle)
|
Nozzle |
Various specification nozzle for option, nozzle inner diameter 2mm、2.5mm、3mm、4mm、6mm |
Power consumption |
35W |
Power supply |
Wide AC90-260V,50Hz/60Hz |
Operation condition |
Temperature 0 to 40 °C, Relative humidity < 80% |
Driver weight |
4.5KG |
IP rate |
IP31 |
[Selection guide]
Filling volume |
Filling time (Sec) |
Accuracy |
Recommend driver |
Recommend pump head |
Recommend tube |
50uL |
≥1.18 |
±3.5% |
BT100-1F |
DG-1A |
0.38×0.85 |
100uL |
≥0.46 |
±2.0% |
1×1 |
||
200uL |
≥0.93 |
±1.2% |
1×1 |
||
0.5mL |
≥0.85 |
±1.2% |
2×1 |
||
1mL |
≥0.80 |
±1.0% |
BT100-1F |
YZ15 |
14# |
2 mL |
≥0.64 |
±1.0% |
16# |
||
3 mL |
≥0.97 |
±1.0% |
16# |
||
5 mL |
≥0.82 |
±1.0% |
25# |
Driver |
Pump head |
Roller of the pump head |
Channel QTY |
Tube |
Speed range (RPM) |
Flow range (mL/min) |
BT100-1F |
DG-1(2,4,6)A |
6 |
1,2,4,6 |
0.6×1 |
0.1-100 |
0.003-3.27 |
1×1 |
0.007-6.45 |
|||||
2×1 |
0.02-20.60 |
|||||
3×1 |
0.04-37.5 |
|||||
DG-1(2,4,6)B |
10 |
1,2,4,6 |
0.6×1 |
0.1-100 |
0.0026-2.64 |
|
1×1 |
0.006-5.96 |
|||||
2×1 |
0.017-16.19 |
|||||
3×1 |
0.028-27.65 |
|||||
BT100-1F |
YZ15 |
3 |
1,2 |
13# |
0.1-100 |
0.008-7.45 |
14# |
0.027-29.87 |
|||||
19# |
0.052-55.7 |
|||||
16# |
0.086-180.4 |
|||||
25# |
0.18-186.6 |
|||||
17# |
0.32-315.0 |
|||||
18# |
0.41-460.0 |
|||||
YZ25 |
3 |
1,2 |
15# |
0.1-100 |
0.17-175.0 |
|
24# |
0.29-350.0 |
|||||
BZ15-14 |
3 |
1,2 |
14# |
0.1-100 |
0.023-24.5 |
|
BZ15-16 |
16# |
0.079-78.5 |
||||
BZ25A |
24# |
0.25-278.4 |
Remarks:
Above data are tested actually by silicone tube, various import tube for option
Import tube to prolong tube lifetime and accuracy
[Dimension]