PRODUCT
Lab syringe pump Current position:Home>Lab syringe pump>Lab syringe pump

LSP01-1B

website:www.landtotech.com  2020-07-18  author:admin    read:

 Product model 

LSP01-1B syringe pump 

[ Production Introduction]  

LSP01-1A is Single channel Syringe Pump, mainly used in laboratory, integrated structure

It suitable for different specification injectors, with high accuracy, small volume liquid transportation

LSP01-1A inner inject function: choose the injector from the list, or input injector inner diameter directly 

Big LCD screen, digital knob combined with membrane keypad, simple and easy operation 

Working mode: Single push

Power-down memory function, after restarting: 

a) Running as before, no need to reset (EEPROM record the parameter)

b) Running according to the preset parameter, or stop(Power-down under flow mode,)

Blocking protection function: when it is blocked, the pump stop working and alert 

Adopt RS485 communication, control through background software 

External port for input/output control function 

Flow calibration function, and get more accurate flow rate

Injection protect function: prevent the injector damaged by adjusting the place of limited block

[Application]

Lab, chemical, electrophysiology, Pharmaceutical, electrostatic spinning

[Technical parameter] 

Operation mode

inject filling

Execution unit number

1

Maximum route distance

140mm

Routing resolution

0.156μm

Liner speed range

5μm/min-65mm/min (flow volume=liner speed x injector inner sectional area)

Liner velocity adjust resolution

5μm/min

Routing control accuracy

deviation≤±0.5% (distance ≥ 30% route )

Rated liner pushing force

90N

Choose injector

main supplier and model no. For choice

Injector defined

 input injector inner diameter directly

Flow calibration

get more accurate flow rate by correcting the program

Operating parameter setting

get more accurate flow rate by correcting the program

Operating parameter setting

liquid volume distribution, injection time etc.

Display parameter selection

user can choose running status after restarting

Status signal output

2 OC door signal output for start/stop/direction

Control signal input

2 OC door signal output for start/stop/direction

Communication port

RS485

Dimensions

280×210×140(mm)

Weight

3.6kg

Power supply

AC 90V-260V/30W

Relative humidity

80%

Working temperature

0-40

 [ Selection guide]

Model

Injector specification

内径(mm)

Injector inner diameter

(μl/min-ml/min)

Flow range reference

LSP01-1A

10μl

0.50

0.001-0.0128

25μl

0.80

0.0025-0.0327

50μl

1.10

0.0048-0.0618

100μl

1.60

0.0101-0.1307

250μl

2.30

0.0208-0.2701

500μl

3.25

0.0415-0.5392

1ml

4.72

0.0875-1.1373

2ml

9.00

0.3181-4.1351

5ml

13.10

0.6739-8.7608

10ml

16.60

1.0821-14.068

20ml

19.00

1.4176-18.429

30ml

23.00

2.0774-27.006

60ml

29.14

3.3346-43.349