PRODUCT
Lab syringe pump Current position:Home>Lab syringe pump>Lab syringe pump

LSP02-1B

website:www.landtotech.com  2020-07-18  author:admin    read:

 Product model 

LSP02-1B Double Channel inject/extract Syringe Pump

[Product Introduction]      

Double channel inject/extract Syringe Pump, mainly used in library, all-in-one structure

Can fill different specification injectors, suitable for high accuracy, small volume liquid transportation

LSP02-1B inner inject function: choose the injector from the list, or input injector inner diameter directly

Big LCD screen, digital knob combined with membrane keypad, simple and fast

4 operation mode: inject, extract, inject and extract, continuous mode

Power-down memory function, after restarting:

a)Running as before, no need to reset (EEPROM record the parameter)

b)Running according to the preset parameter, or stop(Power-down under flow mode,)

Blocking protection function: when it is blocked, the pump stop working and alert

By RS485 communication, control through background software

External port for input/output control function

Calibration function, get more accurate flow rate

Injection protect function: prevent the injector damaged by adjusting the place of limited block

[Application]

Lab, chemical, electrophysiology, Pharmaceutical, electrostatic spinning

[Technical parameter] 

Operation mode

inject, extract

Execution unit number

2

Maximum route distance

140mm

Routing resolution

0.156μm

Liner speed range

5μm/min-65mm/min (flow volume=liner speed x injector inner sectional area)

Liner velocity adjust resolution

5μm/min

Routing control accuracy

deviation≤±0.5% (distance ≥ 30% route )

Rated liner pushing force

180N

Choose injector

main supplier and model no. For choice

Injector defined

 input injector inner diameter directly

Flow calibration

get more accurate flow rate by correcting the program

Operating parameter setting

get more accurate flow rate by correcting the program

Operating parameter setting

liquid volume distribution, injection time etc.

Display parameter selection

user can choose running status after restarting

Status signal output

2 OC door signal output for start/stop/direction

Control signal input

2 OC door signal output for start/stop/direction

Communication port

RS485

Dimensions

280×210×140(mm)

Weight

3.6kg

Power supply

AC 90V-260V/30W

Relative humidity

80%

Working temperature

0-40

[Selection guide]

Model

Injector specification

Injector inner diameter

(mm)

Flow range reference

(μl/min-ml/min)

LSP02-1B

10μl

0.50

0.001-0.0255

25μl

0.80

0.0025-0.0653

50μl

1.10

0.0048-0.1235

100μl

1.60

0.0101-0.2614

250μl

2.30

0.0208-0.5401

500μl

3.25

0.0415-1.0784

1ml

4.72

0.0875-2.2747

2ml

9.00

0.3181-8.2702

5ml

13.10

0.6739-17.522

10ml

16.60

1.0821-28.135

20ml

19.00

1.4176-36.859

30ml

23.00

2.0774-54.012

60ml

29.14

3.3346-86.699